MR quantification of cerebral ventricular volume using a semiautomated algorithm.

L A Johnson 1, J D Pearlman 1, C A Miller 1, T I Young 1, K R Thulborn 1
PMCID: PMC8367507  PMID: 8279334

Abstract

PURPOSE

A semiautomated border identification algorithm, insensitive to user bias, is evaluated for accuracy and speed in the measurement of ventricular volumes from three-dimensional MR images.

METHODS

A three-dimensional gradient-echo technique was implemented on a Signa clinical imaging system. Data from phantoms and patients were analyzed for volume using a segmentation algorithm designed with: 1) correction for partial volume averaging; 2) insensitivity to user bias; and 3) speed. Accuracy, precision, and intra- and interobserver variability were determined.

RESULTS

Average error for phantom studies was 4% to 6%, or 1 to 2 cc across the volumes, which ranged from normal to mild hydrocephalus (< 60 cc). Patient studies showed intra- and interobserver error of 2.3% and 7.8%, respectively. The correction for partial volume averaging resulted in a threefold decrease in error. Data were acquired and reconstructed within 7 minutes. Experienced radiologists required less than 15 minutes to perform each analysis.

CONCLUSIONS

This algorithm allows accurate measurement of ventricular volumes in an efficient, minimally supervised manner.